Martel Electronics BetaGauge 321A User Manual
Page 35
CPRT_COEFB
This command is used for entering a custom RTD into the calibrator. The numeric
value entered after the command will be set as the second coefficient of the
polynomial used by the custom RTD.
For example:
CPRT_COEFB –5.774999E-07 enters –5.774999E-07 as coefficient B.
CPRT_COEFB?
Returns the number, which was entered for the second coefficient of the polynomial
used in the custom RTD. Using the example above, CPRT_COEFB? Would return:
-5.774999E-07
CPRT_COEFC
This command is used for entering a custom RTD into the calibrator. The numeric
value entered after the command will be set as the first coefficient of the polynomial
used by the custom RTD.
For example:
CPRT_COEFC –4.183000E-12 enters –4.183000E-12 as coefficient C.
CPRT_COEFC?
Returns the number that was entered for the third coefficient of the polynomial used
in the custom RTD. Using the example above CPRT_COEFC? Would return:
–4.183000E-12
CPRT_R0
Sets the 0° resistance, R0, in the custom RTD. The value must be entered with a units
label. Refer to the Parameter Units table for assistance.
For example:
CPRT_R0 100 OHM sets R0 to 100 ohms.
CPRT_R0?
Returns the value for the resistance in custom RTD. The above example would return:
1.000000E+02, OHM
DAMP
Turns the dampening function on or off.
For example:
If you send DAMP ON this will turn the dampening function on.
33