Anritsu AWGN MU368060A User Manual
Page 23

3.1
Setting the Parameters
3-5
Example:
Suppose that the following parameters are set for this unit:
Output Level
:
−10 dBm
Digital Mod
: On
System
: Noise1
Noise Bandwidth
: 7.68 MHz
Calculated Bandwidth : 3.84 MHz
From Formula (1),
Power within the calculated bandwidth
= RF output level + Calculated level
= −10 dBm + (−3.45 dB)
= −13.45 dBm
Therefore, the power within the calculated
bandwidth for the RF output is –13.45 dBm.
The power –13.45 dBm is displayed at “(Ab-
solute)”.