IEC 60529 (IP54)
Dust & drip proof
●
Operating System:
Dual integration
●
Display:
Liquid crystal display (maximum count: 1999),
Unit, Mark
Bar graph 30 points max. (20 points on
Ω
range)
●
Over range Indication:
"OL" is shown on the display.
●
Sample Rate:
Approx. 0.5 ~ 2.5 times per second
●
Operating Temperature 0 ~ +40
°
C, relative humidity up to 85%
& Humidity:
●
Storage Temperature
-20 ~ +60
°
C, relative humidity up to 75%
& Humidity:
●
Insulation Resistance:
More than 50M
Ω
at 1000V DC between electrical
circuit and housing case
●
Withstand Voltage:
3700V AC for one minute between electrical
circuit and housing case
●
Overload Protection
Insulation resistance ranges:
1000V Range
1200V (DC+AC p-p) for 10 seconds
500V Range
600V (DC+AC p-p) for 10 seconds
250V Range
300V (DC+AC p-p) for 10 seconds
Continuity ranges:
20/200/2000
Ω
280V (DC+AC p-p) for 10 seconds
Range
(Protection by fusing)
―
7
―
●
Typical Number of Measurements.
(central tendency for supply voltage up to 8V)
Insulation Resistance Ranges:
Approx. 1000 times min. at load 0.5M
Ω
Continuity Ranges:
Approx. 700 times min. at load 1
Ω
●
Operating Errors (IEC 61557-2,-4)
Functions
Range
Measuring range to keep
Maximum percentage
operating error
operating error
20M
Ω
0.50 ~ 19.99M
Ω
1000V
200M
Ω
1.0 ~ 199.9M
Ω
2000M
Ω
10 ~ 1000M
Ω
20M
Ω
0.50 ~ 19.99M
Ω
500V
200M
Ω
1.0 ~ 199.9M
Ω
2000M
Ω
10 ~ 100M
Ω
20M
Ω
0.25 ~ 19.99M
Ω
±
30%
250V
200M
Ω
1.0 ~ 199.9M
Ω
2000M
Ω
10 ~ 100M
Ω
20
Ω
0.20 ~ 19.99
Ω
Ω
200
Ω
1.0 ~ 199.9
Ω
2000
Ω
10 ~ 1999
Ω
The influencing variations used for calculating the operating error are
denoted as follows;
Temperature:
0
℃
and 35
℃
Supply voltage :
8V to 13.8V
●
Applicable Standards
IEC 61010-1
Over voltage CAT III 300V Pollution Degree 2
IEC 61010-2-31
Safety requirements for hand-held probe
assemblies
IEC 61557-1/2/4
Measuring equipment for low voltage
distribution systems
IEC 61326-1
EMC
―