beautypg.com

3 status data, Operation – KROHNE OPTIWAVE 7300C PROFIBUS PA User Manual

Page 8

background image

4

OPERATION

8

OPTIWAVE 7300 C

www.krohne.com

02/2010 - 4000621701 - AD PROFIBUS PA OPTIWAVE 7300

4.1.3 Status data

A measurement value has a 4-byte value and a 1-byte status. Status data is given in the table
that follows:

Status data

CAUTION!
Monitor the status of the device. The device transmits a measurement value when the device
status is "good", "uncertain" or "bad". The status will tell you if the device measures correctly.

Quality

Quality-Substatus

Limits

General

Gr

Gr

QS

QS

QS

QS

Qu

Qu

2

7

2

6

2

5

2

4

2

3

2

2

2

1

2

0

0

0

= bad

0

1

= uncertain

1

0

= good (non Cascade)

Status = bad

0

0

0

0

0

1

= configuration error

0

0

0

0

1

1

= device failure

0

0

0

1

0

0

= sensor failure

0

0

0

1

1

1

= out of service

Status = uncertain

0

1

0

0

0

1

= last usable value

0

1

0

1

1

0

= sub-normal

1

Status = Good / limits

1

0

0

0

1

0

0

0

= Good - no alarm

1

0

0

0

1

0

0

1

= Good - below low limit

1

0

0

0

1

0

1

0

= Good - above high limit

1

0

0

0

1

1

0

1

= Good - Lo-Lo

1

0

0

0

1

1

1

0

= Good - Hi-Hi

1

0

0

0

1

0

1

1

= Constant output

1

0

1

0

0

1

= maintenance required

1 if it is possble that the value is not accurate (for example, the spectrum quality is not sufficient to get a good signal),

this status is shown