KROHNE CORIMASS G-MFM 4085 EEx EN User Manual

Page 9

background image

9

Fig. b Connection without external voltage source (Example). For correct wiring see

tables of input/ output connections


Additional input/output options

No. Option

4*

(Current and
RS485)

Option 5*

(Current and
Modbus)

Option 6

(1 Current,
1 Dual phase
pulse output
and input)

Option C

(2 current,
pulse and
input)

Option D

(3 current

and
pulse)

Option E

(3 current

and
input)

Option F

(3 current

and
alarm

5

Common
(-)

Common
(-)

Common
(-)

Common
(-)

Common
(-)

Common
(-)

Common
(-)

6 Current

output 1 (+)

Current
output 1 (+)

Current
output 1 (+)

Current
output (+)

Current
output 1 (+)

Current
output 1 (+)

Current
output 1 (+)

4

TX/RX

TX/RX

Control Input

Current
output 2 (+)

Current
output 2 (+)

Current
output 2 (+)

Current
output 2 (+)

4.1

TX/RX

TX/RX

Pulse Output
A

Control Input

Current
output 3 (+)

Current
output 3 (+)

Current
output 3 (+)

4.2 +5V

+5V

Pulse

Output

B

Pulse Output

Pulse output

Control Input

Status output
(passive)

The pulse output and status outputs where are passive outputs.
*

Refer to separate RS 485 or Modbus manual



2.3 Electrical connection on converter with intrinsically safe outputs

Note information given on the instrument data plate (voltage,frequency)!


Electrical connection in conformity with IEC 364 or equivalent standard.


The PE ground conductor must be connected to the seperate U-clamp terminal in the
terminal box of the signal converter.


Do not cross or loop the cables in the back of the converter. Please use seperate cable
entries (PG or NPT) for power and signal cables.

Please ensure that the information about power given on the
data plate corresponds to the locally available mains voltage.