KROHNE CORIMASS G-MFM 4085 EEx EN User Manual
Page 9
9
Fig. b Connection without external voltage source (Example). For correct wiring see
tables of input/ output connections
Additional input/output options
No. Option
4*
(Current and
RS485)
Option 5*
(Current and
Modbus)
Option 6
(1 Current,
1 Dual phase
pulse output
and input)
Option C
(2 current,
pulse and
input)
Option D
(3 current
and
pulse)
Option E
(3 current
and
input)
Option F
(3 current
and
alarm
5
Common
(-)
Common
(-)
Common
(-)
Common
(-)
Common
(-)
Common
(-)
Common
(-)
6 Current
output 1 (+)
Current
output 1 (+)
Current
output 1 (+)
Current
output (+)
Current
output 1 (+)
Current
output 1 (+)
Current
output 1 (+)
4
TX/RX
TX/RX
Control Input
Current
output 2 (+)
Current
output 2 (+)
Current
output 2 (+)
Current
output 2 (+)
4.1
TX/RX
TX/RX
Pulse Output
A
Control Input
Current
output 3 (+)
Current
output 3 (+)
Current
output 3 (+)
4.2 +5V
+5V
Pulse
Output
B
Pulse Output
Pulse output
Control Input
Status output
(passive)
The pulse output and status outputs where are passive outputs.
*
Refer to separate RS 485 or Modbus manual
2.3 Electrical connection on converter with intrinsically safe outputs
–
Note information given on the instrument data plate (voltage,frequency)!
–
Electrical connection in conformity with IEC 364 or equivalent standard.
–
The PE ground conductor must be connected to the seperate U-clamp terminal in the
terminal box of the signal converter.
–
Do not cross or loop the cables in the back of the converter. Please use seperate cable
entries (PG or NPT) for power and signal cables.
Please ensure that the information about power given on the
data plate corresponds to the locally available mains voltage.