DEVICE DESCRIPTION
22
OPTIMASS 3000
www.krohne.com
03/2012 - 4001593001 - MA OPTIMASS 3000 R01 en
On all meters operating on gas measurement the casing of the meter is fitted with a burst disc. If
the primary seal (tube/s) fails leakage will occur from the burst disc. Install the meter so that the
burst disc is pointing away from personnel.
Regular maintenance of the burst disc:
Regular maintenance of the burst disc:
Regular maintenance of the burst disc:
Regular maintenance of the burst disc:
Carry out regular maintenance checks on burst discs for leakage and/or blockages. On all
OPTIMASS meters, the primary seal is considered to be the measuring tube of the meter. The
materials of construction of the measuring tube/s are described within the relevant sections of
this handbook and the customer’s product and any other fluid flowing through the tube must be
compatible with the material of construction. If failure of the primary seal is suspected then the
process line should be de-pressurised and the meter removed as soon as it is safe to do so.
Please contact customer service for servicing or replacement of the meter.
3.4 Temperature differential and thermal shock
Temperature differential
Temperature differential
Temperature differential
Temperature differential
The maximum difference between ambient temperature and process (operating) temperature is
110°C / 230°F.
Thermal shock
Thermal shock
Thermal shock
Thermal shock
Thermal shock occurs when there is a sudden and extreme change (shift) in process
temperature. To avoid thermal shock, refer to the following table for the maximum temperature
shift.
Meter
Meter
Meter
Meter
Maximum temperature shift
Maximum temperature shift
Maximum temperature shift
Maximum temperature shift
Stainless Steel
+80°C / +176°F
Hastelloy®
+80°C / +176°F
CAUTION!
Operation outside these limits may result in shifts in density and mass flow calibration. Repeated
shocking may also lead to premature failure of the meter! However, higher thermal shocks are
possible at lower working pressures. For more information, please contact your nearest
representative.
MA_OPTIMASS_3000_en_120315_4001593001_R01.book Page 22 Tuesday, October 23, 2012 10:14 AM