2

INSTALLATION

14

OPTIMASS 3000

www.krohne.com

03/2012 - 4001593001 - MA OPTIMASS 3000 R01 en

2.4.3 Self draining

Verticle mounting angles for self draining

Verticle mounting angles for self draining

Verticle mounting angles for self draining

Verticle mounting angles for self draining

If the meter is mounted vertically and the installation requires the meter to be self draining,
mount the meter at the angle shown in the table. The angles are marked on the meter base
plate.

Set angles

Figure 2-3: Angle of rotation for self draining

1 7º of clockwise rotation from the vertical for self draining. (See table for applicable meter sizes).
2 13º of clockwise rotation from the vertical for self draining. (See table for applicable meter sizes).

Meter size

Meter size

Meter size

Meter size

Angle of rotation (clockwise)

Angle of rotation (clockwise)

Angle of rotation (clockwise)

Angle of rotation (clockwise)

01

03

13º

04

13º