TECHNICAL DATA
34
DK32 - DK34 - DK37
www.krohne.com
06/2014 - 4000287103 MA DK32-34-37-R05-en
7.4 Flow table
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:
The flow measurement of gases are refered to
Nl/h or Nm
3
/h: Volume flow in normal state 0°C, 1.013 bar abs. (DIN 1343)
SCFM or SCFH: Volume flow in standard state 15°C, 1.013 bar abs. (ISO 13443)
Measuring span:
10 : 1
Declaration of flow:
Values = 100%
Water: 20°C [68°F]
Air: 20°C [68°F], 1.013 bara [14.7 psia]
Water flow rate
Air flow rate
Pressure drop
Cones
[l/h]
[GPH]
[Nl/h]
[SCFH]
[mbar]
[psig]
K 005
-
-
16
1
0.6
1
14
0.21
K 005
-
-
50
1.9
31
0.46
K 010
1.5
1
0.40
1
70
1
2.6
1
66
0.97
K 010
3
0.8
100
3.7
66
0.97
K 015
5
1.3
150
5.6
19
0.28
K 040
10
2.5
400
15
27
0.40
K 080
25
6.5
800
30
55
0.81
K 125
40
11
1250
45
42
0.62
K 200
60
16
2000
75
85
1.25
K 300
80
20
2500
90
117
1.72
K 340
100
25
3400
130
166
2.44
1 with titanium float
INFORMATION!
The operating pressure should be at least twice the pressure loss for liquids, and at least 5
times the pressure loss for gases! The specified pressure drops are valid for water and air at
maximum flow rate. Other flow ranges on request. Conversion of other media or operating data
(pressure, temperature, density, viscosity) is performed using the calculation method in
accordance with VDI /VDE Directive 3513.