5

START-UP

22

DK32 - DK34 - DK37

www.krohne.com

06/2014 - 4000287103 MA DK32-34-37-R05-en

Start-up

5.1 Standard device

CAUTION!

When starting up the device, the following points must be observed:

Compare the actual operating pressure and the product temperature of the system with the
specifications on the nameplate (PS and TS). These specifications may not be exceeded.

Make sure materials are compatible.

Slowly open the shut-off valve.

When measuring liquids, vent the pipes carefully.

When measuring gases, increase pressure slowly.

Avoid float impact (e.g. caused by solenoid valves), as this is likely to damage the measuring
unit or float.