background image

User Guide for Optiflex (TDR) DD of AMS

Dev Rev-01, DD Rev-02



© 2006 KROHNE SAS 26103 Romans (France)

Page

10

of

42

3.4.2 Output1

Upper


In Output1 Upper, user allows to trim current output1.
The user start to fix the current to 20mA, after entering the current value he read on a voltmeter,
this value is sent to the device. This method allows adjusting to have exactly the good current
value of output1 loop.