2.8 Electrical data

The electronic signal output may only be connected to intrinsically safe circuits. Depending on
the instrument design, the following maximum values apply per circuit:

Design DK3./../.././../A–Ex

Irrespective of the instrument design the following values are to be observed for each
intrinsically safe circuit in case of interconnection:

U

i

16 VDC

I

i

25 mA

P

i

64 mW

C

i

150 nF

L

i

150 µH

AD_DK32_34_NEPSI_Ex_i_R01_en_4002598201_PRT.book Page 7 Friday, March 1, 2013 2:43 PM