2.7 Ambient temperature / temperature classes

Due to the influence of the product temperature, variable area flowmeters with built-in electrical
equipment (electric variants) are not assigned to any fixed temperature class. The temperature
class of these devices is rather a function of the product temperature and ambient temperature
that is present and the specific device version. Please see the following tables for the
assignments.

The tables take into account the following parameters:

• Ambient temperature T

amb.

• Product temperature T

m

DK3./../.././../A–Ex permissible medium and ambient temperatures

INFORMATION!

The maximum permissible product temperatures listed in the tables are valid under
the following conditions:

The measuring device is installed and operated in accordance with the installation
instructions in the installation and operating manual.

It must be ensured that the flowmeter is not heated by the effects of additional
heat radiation (sunshine, neighbouring system components) and
thus operated above the permissible ambient temperature range.

Insulation must be limited to the piping.
Unobstructed ventilation of the indicator part must be ensured.

Temperature
class

Ambient
temperature

Maximum permissible medium temperature
with connector (S) or cable gland (L)

Type DK32

Type DK34

[°C]

[°F]

[°C]

[°F]

[°C]

[°F]

T6

-20...+40

-4...+104

75

167

80

176

-20...+50

-4...+122

70

158

70

158

-20...+60

-4...+140

60

140

60

140

T5

-20...+40

-4...+104

100

212

100

212

-20...+50

-4...+122

95

203

100

212

-20...+60

-4...+140

85

185

90

194

T4

-20...+40

-4...+104

135

275

135

275

-20...+50

-4...+122

130

266

135

275

-20...+60

-4...+140

120

248

130

266

-20...+90

-4...+194

90

194

90

194

T3...T1

-20...+40

-4...+104

135

275

150

302

-20...+50

-4...+122

130

266

140

284

-20...+60

-4...+140

120

248

130

266

-20...+90

-4...+194

90

194

90

194