7 CONNECTION DIAGRAM

Flow tube

Hazardous locations

of Zone 1 and 2

E

IFS 4000…-EEx

Primary Head

Intrinsically safe electrode circuits

Increased safe field coil circuit

Coil

B1 B

B2 I+ I

L N

BINARY

CURRENT

MAINS

OUTPUTS

OUTPUT

SUPPLY

TERMINAL COMPARTMENT

Standard "EEx e" (Optional "EEx d")

ELECTRONICS COMPARTMENT (always "EEx d")

Intrinsically safe ("ib")

Increased safe ("e")

electrode circuits

field coil circuits

(No. "2", "3", "1")

(No. "7", "8")

IFC 090-EEx

Signal Converter

PE
FE

L

N

PE 100-230 Vac

SIGNAL IN-/OUTPUTS

FE 24 Vac/dc

EQUIPOTENTIAL BONDING CONDUCTOR

4 mm

2

(OPTIONAL)

Flameproof (EEx d) cable

feed-through LC-2/EEx

Field coil wires - green/blue

(PTFE insulated copper)

B

A

E

Coil

Electrode cables - white/pink

(PTFE insulated shielded copper)

IFC 090i-EEx

Signal Converter

x

x

x

x

1L½ 0L½

PE
FE

INTRINSICALLY SAFE
SIGNAL IN-/OUTPUTS

L

N

PE

100-230 Vac

(i.e. MODIS)

FE

24 Vac/dc

Separation plate

Unused terminal

OPTION: MODIS

Flameproof (EEx d)

terminal feed-through

B

A