3

ELECTRICAL CONNECTIONS

18

TIDALFLUX 4300 F

www.krohne.com

09/2010 - 7308302300 - QS TIDALFLUX 4000 R03 en

1 Strip the conductor to dimension a.
2 Trim the outer shield (60) to dimension b and pull it over the outer sheath.
3 Remove the stranded drain wire (6) of the outer shield and the shields and stranded drain

wires of the insulated conductors (2, 3). Remove the inner shield (10). Be sure not to damage
the stranded drain wire (1).

4 Slide an insulating tube over the stranded drain wire (1).
5 Crimp the wire end ferrules onto conductors 2 and 3 and the stranded drain wire (1).
6 Pull the heat-shrinkable tubing over the prepared signal cable.

Figure 3-7: Preparing signal cable B, connection to measuring sensor

a = 50 mm / 2"
b = 10 mm / 0.39"