Technical data – KROHNE SMARTSENS ORP 8510 EN User Manual
Page 26
6
TECHNICAL DATA
26
SMARTSENS ORP 8510
www.krohne.com
11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en
Communication
ORP range
-1500...+1500 mV
Output signal
4...20 mA (passive)
Output resolution
20 µA
Field communication
HART
®
7 - FSK 1200 physical layer definition on top of the current loop
Filter adjustable
1...60 seconds
Electrical connections
Power supply
15...30 VDC loop powered
Measuring range
4...20 mA + HART
®
protocol
Load
Minimum 0 Ω; maximum R
L
= ((U
ext.
- 15 VDC) / 22 mA)
Error signal
Acc. to NAMUR NE 43
Upper value: ≥ 21.0 mA
HART
®
HART
®
protocol via current output
Device revision
1
Physical layer
FSK
Device category
Sensor, galvanically isolated
System requirements
250 Ω loop resistance for HART
®
communication
Multidrop operation
4 mA
Approvals
CE
This device fulfils the statutory requirements of the EC directives. The
manufacturer certifies successful testing of the product by applying the CE
mark.
Shock resistance:
IEC 60068-2-31, Environmental testing – Part 2: Test Ec
Electromagnetic compatibility:
Acc. to EN 61326, NAMUR NE 21
Low voltage directive:
Safety requirements for electrical equipment for measurement, control, and
laboratory use in accordance with EN 61010-1:2001
INFORMATION!
For further information contact your local sales office.