OPERATION

4

19

SMARTSENS ORP 8510

www.krohne.com

11/2013 - 4002547502 - MA SMARTSENS ORP 8510 R02 en

• Press Enter

Enter

Enter

Enter to proceed.

• Select no

no

no

no to deactivate the function manual hold

manual hold

manual hold

manual hold.

• Press Enter

Enter

Enter

Enter to leave the calibration menu.

4.2 Troubleshooting

CAUTION!
Make sure that the sensor tip is clean and dust-free. If necessary, clean the tip as described on
page 21.

CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and
deliver erratic readings!
If moisture has entered the connector dry it with air (e.g. hair blower).

Problem

Possible cause

Remedy

The ORP sensor

does not deliver a

signal.

Mechanical damage of the

glass, e.g. small cracks.

Exchange sensor.

Open circuit

Check the cable wiring of the sensor cable on the junction box.

Otherwise connect the sensor to the primary master e.g.
PACTware

TM

FTD/DTM or to the HART

®

handheld to exclude a sensor

damage.

The ORP sensor

delivers an

unstable signal.

The diaphragm in the

reference half cell does

not provide good contact

to the process medium

due to drying up or

coatings.

Clean the diaphragm with hot soap or acid using a soft tissue
(details on page 21).

Submerge sensor in water and increase the temperature to
+50...+60°C / +122...+140°F.

Submerge sensor in 3 molar KCl solution at ambient temperature.
The decrease in temperature will cause the reference half cell to
suck in KCl solution through the diaphragm and regenerate the
diaphragms functionality.