beautypg.com

Technical data, 2 technical data – KROHNE SMARTSENS ORP 1590 EN User Manual

Page 24

background image

6

TECHNICAL DATA

24

SMARTSENS ORP 1590

www.krohne.com

11/2013 - 4002547102 - MA SMARTSENS ORP 1590 R02 en

6.2 Technical data

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

Measuring system

Measuring principle

Potentiometric

Measuring range

-1500...+1500 mV

Design

Shaft diameter

20 mm / 0.78"

Insertion length

23 mm / 0.91"

Temperature sensor

Pt1000

Connector

VarioPin 2.0 (VP 2.0)

Operating conditions

Temperature range

0...+80°C / +32...+176°F

Process pressure

6.9 bar at +60°C / 100 psi at +140°F

Measuring accuracy

0.5%

Conductivity

Min. 150 µS/cm

Installation conditions

Ingress protection

IP 68

Weight

Approx. 193 g / 0.43 lb

Process connection

3/4-14 NPT (male)

Materials

Sensor shaft

CPVC

Measuring electrode

Platinum

Inner buffer

pH 7.0

Reference

KCl gel double junction

Diaphragm

Ceramic

Sensor head

Nickel plated brass body with VP 2.0 connector

.book Page 24 Wednesday, October 30, 2013 11:26 AM