background image

4

OPERATION

18

SMARTSENS ORP 1590

www.krohne.com

11/2013 - 4002547102 - MA SMARTSENS ORP 1590 R02 en

4.1.2 Calibration with HART

®

Handheld 475 FIELD COMMUNICATOR

Manual hold

• Start the function calibration

calibration

calibration

calibration in menu mode Quick Setup

Quick Setup

Quick Setup

Quick Setup or Setup

Setup

Setup

Setup.

• Activate the function manual hold

manual hold

manual hold

manual hold to avoid an alarm.

• Select Yes

Yes

Yes

Yes.

• Press Enter

Enter

Enter

Enter to proceed.

Start calibration procedure

• Set value of ORP solution; Default setting: 468 mV. Press the key >>>> to enter the submenu. Now

you can change the default setting.

• Press Enter

Enter

Enter

Enter and Ok

Ok

Ok

Ok to proceed.

• Dip the sensor into the ORP solution and wait till the value is stable.
• Press Ok

Ok

Ok

Ok to proceed.

i

The measurement of ORP solution starts. After approx.10 seconds and a successful
calibration the following values appear:
Old ORP value in mV
ORP solution in mV
Electrode voltage in mV
New offset in mV
New ORP value in mV

• Press Ok

Ok

Ok

Ok to proceed.

• Set the calibration date with DD-MM-YYYY. Press the key >>>> to enter the submenu. Now you can

change the calibration date .

• Press Enter

Enter

Enter

Enter and Ok

Ok

Ok

Ok to proceed.

i

The message Save values?

Save values?

Save values?

Save values? appear.

• Press yes

yes

yes

yes to save the values.

• Press Enter

Enter

Enter

Enter to proceed.

• Select no

no

no

no to deactivate the function manual hold.

• Press Enter

Enter

Enter

Enter to leave the calibration menu.

A ORP calibration is necessary in regular intervals or when installing a new ORP sensor.

If the calibration was not successful, the message Warning!!!

Warning!!!

Warning!!!

Warning!!! appears behind the value New

New

New

New

offset

offset

offset

offset.

• Press Ok

Ok

Ok

Ok to proceed.

• Set the calibration date with DD-MM-YYYY. Press the key >>>> to enter the submenu. Now you can

change the calibration date .

• Press Enter

Enter

Enter

Enter and Ok

Ok

Ok

Ok to proceed.

i

The messages Warning!!! Offset out of range!

Warning!!! Offset out of range!

Warning!!! Offset out of range!

Warning!!! Offset out of range! and Save values?

Save values?

Save values?

Save values? appear.

• Select betweenyes

yes

yes

yes or no

no

no

no.

CAUTION!
Make sure that the sensor tip is clean and dust-free. If necessary, clean the tip as described on
page 20.

CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and
deliver erratic readings!
If moisture has entered the connector dry it with air (e.g. hair blower).