Operation – KROHNE SMARTSENS ORP 1590 EN User Manual
Page 19
OPERATION
4
19
SMARTSENS ORP 1590
www.krohne.com
11/2013 - 4002547102 - MA SMARTSENS ORP 1590 R02 en
• Press Enter
Enter
Enter
Enter to proceed.
• Select no
no
no
no to deactivate the function manual hold
manual hold
manual hold
manual hold.
• Press Enter
Enter
Enter
Enter to leave the calibration menu.
4.2 Troubleshooting
4.3 Status messages and diagnostic information
Measurements out of specification
Maintenance
CAUTION!
Make sure that the sensor tip is clean and dust-free. If necessary, clean the tip as described on
page 20.
CAUTION!
Moisture inside the sensor connector must be avoided! Moisture may cause a shortcut and
deliver erratic readings!
If moisture has entered the connector dry it with air (e.g. hair blower).
Problem
Possible cause
Remedy
The ORP sensor
does not deliver a
signal.
Mechanical damage of the
glass, e.g. small cracks.
Exchange sensor.
Open circuit
Check the cable wiring of the sensor cable on the junction box.
Otherwise connect the sensor to the primary master e.g.
PACTware
TM
FTD/DTM or to the HART
®
handheld to exclude a sensor
damage.
The ORP sensor
delivers an
unstable signal.
The diaphragm in the
reference half cell does
not provide good contact
to the process medium
due to drying up or
coatings.
•
Clean the diaphragm with hot soap or acid using a soft tissue
(details on page 20).
•
Submerge sensor in water and increase the temperature to
+50...+60°C / +122...+140°F.
•
Submerge sensor in 3 molar KCl solution at ambient temperature.
The decrease in temperature will cause the reference half cell to
suck in KCl solution through the diaphragm and regenerate the
diaphragms functionality.
Message
Message
Message
Message
Description
Description
Description
Description
Action
Action
Action
Action
S01
ORP value > ORP max
ORP value out of range
Consider the measuring range limits, otherwise
select a suitable sensor for the process conditions of
the application.
S02
ORP value < ORP min
ORP value out of range
S03
Temp. value > Temp max
Temperature value out of range
S04
Temp. value < Temp min
Temperature value out of range
Message
Message
Message
Message
Description
Description
Description
Description
Action
Action
Action
Action
M01
Offset < -99 mV
Recalibrate the sensor. For more information refer
to
Calibration with PACTwareTM on page 17
M02
Offset> +99 mV
M03
Maintenance interval expired