Step 2: Configure /adjust the temperature sensor
• If necessary, enter the temperature correction in Kelvin so that the signal converter shows the
same temperature as the reference thermometer. Press ^ to confirm the entered value. The
temperature sensor has been adjusted.
Step 2a: probe Pt100/1000
probe Pt100/1000
probe Pt100/1000
probe Pt100/1000
Read the currently measured temperature of the Pt100/1000 temperature sensor from the measurement
screen and write it down.
Measure the temperature with a reference thermometer and check if it deviates from the temperature
measured by the Pt100/1000.
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu setup
setup
setup
setup is highlighted.
•
Press
or
until the main menu setup
setup
setup
setup is highlighted.
MAIN MENU
A quick setup
B test
> C setup
> C setup
> C setup
> C setup
D service
•
Press >>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "setup" and "c1" appears,
beneath the submenu process input A
process input A
process input A
process input A is highlighted. Choose process input A or B where ever
pH is configurated.
•
Press
or
to select process input A
process input A
process input A
process input A or process input B
process input B
process input B
process input B.
•
Press >>>> to enter the chosen menu process input A/B
process input A/B
process input A/B
process input A/B.
You are on the second submenu level. Press or until the submenu temperature
temperature
temperature
temperature is
highlighted.
•
Press >>>> to enter the chosen menu. The submenu probe Pt100/1000
probe Pt100/1000
probe Pt100/1000
probe Pt100/1000 is highlighted.
•
Press
or
until the submenu correction
correction
correction
correction is highlighted.
•
Press ^ to enter the chosen menu