OPERATION

22

OPTISENS PH 8390

www.krohne.com

03/2012 - 4001930601 MA OPTISENS PH 8390 R01 en

4.2.2 Menu B, test

The procedure to start the simulation process is the same for all functions:

• Choose the function with the help of ↓ or ↑ and press ^.

i

You see the two options "set value" (opens the editor to enter the simulation value) and
"break" (exits the menu without simulation).

• Choose the desired option with the help of ↑ or ↓ and press ^.

i

If you chose "set value", the device asks "start simulation" and offers the options "no" (exits
the menu without simulation) or "yes" (starts the simulation finally).

• Choose the desired option with the help of ↑ or ↓ and press ^.

i

If you chose "yes", the simulation starts.

B1,sim.process inp.A
B2,sim.process inp.B

B4, actual values

INFORMATION!
Note that the appearance of some sub-menus depends on the hardware setting and the used
sensor(s). Also only the sensor relevant menus and sub-menus are shown here in detail. For all
other menu functions refer to the MAC 100 signal converter manual.

Level

Designation / function

Settings / descriptions

B1.1

B1.1

B1.1

B1.1

B2.1

B2.1

B2.1

B2.1

temperature

In this menu the temperature can be simulated.

B1.5

B1.5

B1.5

B1.5

B2.5

B2.5

B2.5

B2.5

REDOX/ORP

ONLY if switched to ORP!

ONLY if switched to ORP!

ONLY if switched to ORP!

ONLY if switched to ORP!

(for dual channel version

only; refer to pH/ORP

sensor manual for further

information)

In this menu the relative concentration of Redox/ORP can be simulated. For

dual channel version only; refer to pH/ORP sensor manual for further

information. ONLY if switched to ORP!

B1.7

B1.7

B1.7

B1.7

B2.7

B2.7

B2.7

B2.7

pH

(for dual channel version

only; refer to pH/ORP

sensor manual for further

information)

In this menu the relative concentration of pH can be simulated. For dual

channel version only; refer to pH/ORP sensor manual for further

information.

Level

Designation / function

Settings / descriptions

This menu groups several functions which allow to display the corresponding actual reading. The shown measurements

are depending on the device configuration.

B4.1

B4.1

B4.1

B4.1

operating hours

This menu shows the operating time of the devices in hours.

B4.2

B4.2

B4.2

B4.2

process input A

In this menu the measurements from process input A can be read.

B4.3

B4.3

B4.3

B4.3

process input B

In this menu the measurements from process input B can be read.For dual

channel version only; refer to pH/ORP sensor manual for further

information.