Operation – KROHNE OPTISENS PH 8390 EN User Manual
Page 24
4
OPERATION
24
OPTISENS PH 8390
www.krohne.com
03/2012 - 4001930601 MA OPTISENS PH 8390 R01 en
C1.15.5
C2.15.5
temp. comp.
Menu for activating the temperature compensation parameters for the
measurement.
Options:
•
Linear: linear temperature compensation.
•
Off: temperature compensation is disabled.
C1.16
C1.16
C1.16
C1.16
C2.16
C2.16
C2.16
C2.16
REDOX cal.
menu item for calibrating the ORP sensor
C1.16.1
C2.16.1
prepare calibration
view actual redox value
C1.16.2
C2.16.2
ref. value Redox
Enter the reference value of the probe in mV
C1.16.3
C2.16.3
start calibration
Start calibration procedure.
C1.16.4
C2.16.4
stored value
View stored value of calibration.
C1.17
C1.17
C1.17
C1.17
C2.17
C2.17
C2.17
C2.17
pH cal.
menu item for calibration the pH sensor
C1.17.1
C2.17.1
temp.comp
Menu for activating the temperature compensation parameters for the
calibration
Options:
•
off: temperature measurement is disabled
•
manual: temperature value has to be entered manually
•
automatic: temperature measurement is performed as configured
C1.17.2
C2.17.2
start calibration in buffer
1
Start calibration procedure.
C1.17.3
C2.17.3
stored value
View stored value of calibration.
C1.17.4
C2.17.4
reference
Enter the reference value of the probe in pH
C1.17.6
C2.17.6
start calibration in buffer
2
Start calibration procedure.
C1.17.7
C2.17.7
stored value
View stored value of calibration
C1.17.8
C2.17.8
reference
Enter the reference value of the probe in pH
CAUTION!
If you choose for measurement the temperature compensation "linear" than choose between
"automatic" or "manual" for the temperature compensation during calibration. If you choose for
measurement the temperature compensation "off" than choose also "off" for the temperature
compensation during calibration.
Level
Designation / function
Settings / descriptions
.book Page 24 Monday, March 19, 2012 1:52 PM