7

TECHNICAL DATA

48

OPTITEMP TRA/TCA

www.krohne.com

08/2010 - 4000630501 - MA OPTITEMP TRA/TCA R01 en

7.3.4 Measuring inserts: Diameter and standard lengths

The diameter the thermometer's measuring insert must have depends on the existence of a
thermowell and its tip:

• Thermowell with reduced tip: Measuring insert with Ø 3 mm / 0.12"
• Straight thermowell and thermowell with tapered tip: Ø 6 mm / 0.24"
• Version without thermowell: Ø 6 mm / 0.24"

Insertion-type thermometer (TRA/TCA-P10/20/40)

Threaded thermometer (TRA/TCA-S11/21)

INFORMATION!
The lengths given for the thermowells and measuring inserts in the following tables are
standard lengths. The manufacturer can provide other lengths on request.

Thermowell length

Length of the measuring insert

[mm]

["]

[mm]

["]

305

12.01

315

12.40

395

15.55

405

15.94

545

21.46

555

21.85

INFORMATION!
The following applies when determining the length of the measuring insert for insertion-type
thermometers: length of the thermowell + 10 mm / 0.4".

Insertion length

Length of the measuring insert

[mm]

["]

[mm]

["]

75

2.95

120

4.7

100

3.94

145

5.7

115

4.53

160

6.3

270

10.63

315

12.4

390

15.35

435

17.1

INFORMATION!
The following applies when determining the length of the measuring insert for threaded
thermometers: Distance from the underside of the seal (G thread) / two-thirds of the thread
height from the bottom (NPT thread) to the tip of the thermowell + 45 mm / 1.8".