Installation site, angle and length

These three parameters depend upon the space available and the diameter of the pipe. The
manufacturer generally recommends three possible installation scenarios in pipes with flowing
media:

• Small pipe diameter: Installation directly against the direction of flow in a bend in the

pipe (1).

• Small pipe diameter: Installation diagonally against the direction of flow, if a bend in the pipe

is available (2).

• Large pipe diameter: Vertical installation, if flow-induced periodic vortex shedding does not

cause the thermometer to vibrate in its resonance frequency (3).