beautypg.com

6 set up, 1 indication of the switching status, 2 function test – KROHNE OPTISWITCH 4000C Contactless EN User Manual

Page 18

background image

6 Set up

6.1 Indication of the switching status

The switching status of the electronics can be checked on the

upper part of the housing.

6.2 Function test

OPTISWITCH 4000 C has an integrated test switch which can

be activated magnetically. Proceed as follows to test the

instrument:

à

Hold the test magnet (accessory) to the magnet symbol on

the instrument housing

Fig. 11: Function test

The test magnet changes the current switching condition of the

instrument. You can check the change on the signal lamp.

Please note that the connected instruments are activated

during the test.
If OPTISWITCH 4000 C does not switch over after several

tests with the test magnet, you have to check the plug

connection and the connection cable and test the instrument

again. If there is no switching function, the electronics will be

defective. In this case you have to exchange the electronics or

return the instrument to our repair department.