beautypg.com

5 connecting to power supply, 1 preparing the connection – KROHNE OPTISOUND 3010C 2wire EN User Manual

Page 17

background image

5 Connecting to power supply

5.1 Preparing the connection

Always observe the following safety instructions:

l

Connect only in the complete absence of line voltage

l

If overvoltages are expected, overvoltage arresters should

be installed.

In hazardous areas you should take note of the appropriate

regulations, conformity and type approval certificates of the

sensors and power supply units.

Power supply and current signal are transmitted via the same

two-wire connection cable. The power supply range can differ

depending on the instrument version. The exact range is

stated in the "Technical data" in the Supplement.
Provide a reliable separation between the supply circuit and

the mains circuits acc. to DIN VDE 0106 part 101.
Bear in mind the following factors regarding supply voltage:

l

the reduction of the output voltage of the power supply unit

under nominal load (with a sensor current of 20.5 mA or

22 mA in case of fault signal)

l

the influence of additional instruments in the circuit (see

load values in Technical data).

OPTISOUND 3010 C is connected with standard two-wire

cable without screen. An outer cable diameter of 5 ... 9 mm

ensures the seal effect of the cable entry. If electromagnetic

interference is expected, we recommend the use of screened

cable.

Connect the cable screen on both ends to ground potential. In

the sensor, the screen must be connected directly to the

internal ground terminal. The ground terminal outside on the

housing must be connected to the potential equalisation (low

impedance).
If potential equalisation currents are expected, the connection

on the evaluation side must be made via a ceramic capacitor

(e.g. 1 nF, 1500 V). The low frequency potential equalisation

currents are thus suppressed, but the protective effect against

high frequency interference signals remains.

Note safety instructions

Take note of safety

instructions for Ex

applications

Select power supply

Selecting the connection ca-

ble

Cable screening and ground-

ing

OPTISOUND 3010 C - 4 … 20 mA/HART

17

Connecting to power supply

30503

-EN

-050914