5.5 Error messages

5.5.1 Error messages in measurement mode

Error message
(display 2

nd

line)

Type Description

Corrective

action

required

INTL.ERR.nn

*

Internal error in converter
operation

Switch off the power and try again. If the problems
persists contact KHRONE MARSHALL service.

NO SIGNAL

N

No signal from the vortex
sensor

No flow through the primary or Vortex sensor
problem

CHECK INST

N

Vortex sensor signal quality
is bad

Check that 1. Flow rate > q min if OK. 2. Check for
excessive pipe vibration & proper installation.

LOW SIGNAL

N

Vortex signal amplitude too
low.

Check that 1. Flow rate > q min if OK contact
KHRONE MARSHALL Service.

HIGH SIGNAL

N Vortex sensor signal

amplitude too high

This occurs very rarely. Check 1. Flow rate < q max
if OK contact KHRONE MARSHALL Service.

LOW.FREQ.

N

Vortex frequency too low

Check 1. Flow rate > q min else call KHRONE
MARSHALL Service.

HIGH.FREQ.

N

Vortex frequency too high

check 1. Flow rate < q max else call KHRONE
MARSHALL Service.

LOW.FLOW

N Flow rate lower than

minimum flow rate q min.

Converter will continue to display actual flow rate.
However, accuracy of measurement may suffer. If
flow rate reduces further, then other errors such as
CHECK INSTALL,LOW SIGNAL will be generated

HIGH.FLOW

N

Actual flow rate higher than
q max.

Corrective action depends on application process. If
flow rate exceeds too much further it may damage
entire instrument physically !

LOW.TEMP. OPR

N

Operating temp. lower than
limit.

Corrective action depends on the process.

HIGH.TEMP.OPR

N

Operating temp. higher than
limit

Corrective action depends on the process.

LOW.PRES. OPR

N Operating pressure lower

than limit.

Take corrective action depending on the process.

HIGH.PRES. OPR

N

Operating pressure higher
than limit.

Corrective action is process dependent.

LOW.TEMP. PHY.

N

Operating temp. lower than
physical limit.

Take corrective action depending on the process.

HIGH.TEMP.PHY

F

Operating temp. higher than
physical limit.

Will cause damage to primary as well as signal
converter !

LOW.PRES.PHY.

N Operating pressure lower

than physical limit.

Take corrective action depending on process.

HI.PRES.PHY

F

Operating pressure higher
than physical limit.

Will cause damage to primary and signal converter!

T.SENS.SHORT

N

Temp. sensor / wires short
circuit.

T.SENS.OPEN

N

Temp. sensor open circuit.

P.SENS.OPEN

N Pressure sensor open

circuited


Indicates fault in temperature/pressure sensor.
Contact KHRONE MARSHALL Service. You may
use signal converter with temperature or pressure
compensation offline in the mean time till you get
the service. P&T values entered through menu are
used.

INV. CONFIG

F

Configuration data in non-
volatile memory is not valid.

Check entire configuration again. If error persists-
call for service