beautypg.com

Electrical connections, 3 "ex i" i/o connections – KROHNE OPTISONIC 6300 EEx EN User Manual

Page 19

background image

ELECTRICAL CONNECTIONS

4

19

OPTISONIC 6300

www.krohne.com

03/2010 - 7309972300 - EX OPTISONIC 6300 R03 en

4.3 "Ex i" I/O connections

The following intrinsically safe I/O connections are available:

The I/O circuits titled “Exi-IO” and “Exi-Option” are always provided with type of protection
Intrinsic Safety (Ex ia).

Up to a maximum of 4 intrinsically safe (Ex ia) in-/outputs are possible. All intrinsically safe
circuits are galvanically isolated with respect to earth and each other. To avoid summation of
voltages and current, the wiring of these “Ex ia”-circuits must be sufficiently separated, e.g. in
accordance with the requirements of standard EN 60079-14, clause 12.2.
The “Ex ia” in-/outputs may only be connected to other “Ex ia” or “Ex ib” approved devices (e.g.
intrinsically safe isolation amplifiers), even if such devices are installed in a non-hazardous
location !
Connection to non-“Ex i” devices cancels the “Ex ia” properties of the flowmeter.

Terminals L and N (or L+ and L-) for connection of the mains supply are not available with type of
protection “intrinsic safety”. To achieve the necessary separation distances according to EN
60079-11 between the non-”Ex i” and “Ex i” circuits, the mains terminals are provided with a
semi-circular protection cover with a “snap-in” lock. This cover MUST be closed before
establishing the power supply to the converter.

I/O PCB

I/O PCB

I/O PCB

I/O PCB

I/O functions

I/O functions

I/O functions

I/O functions

Exi-IO

Current Output + HART

communication

Pulse / Status Output

Current Output, active + HART

communication

Ex ia IIC

U

i

= 30 V, I

i

= 100 mA, P

i

= 1,0 W

C

i

= 10 nF, L

i

= negligibly low

Ex ia IIC

U

o

= 21 V, I

o

= 90 mA, P

o

= 0,5 W

Lineair characteristics

C

o

= 90 nF, L

o

= 2,0 mH

C

o

= 110 nF, L

o

= 0,5 mH

Exi-Option

Current Output

Pulse / Status Output / Control

Input

Current Output, active

Ex ia IIC

U

i

= 30 V, I

i

= 100 mA, P

i

= 1,0 W

C

i

= 10 nF, L

i

= negligibly low

Ex ia IIC

U

o

= 21 V, I

o

= 90 mA, P

o

= 0,5 W

Lineair characteristics

C

o

= 90 nF, L

o

= 2,0 mH

C

o

= 110 nF, L

o

= 0,5 mH