2.1 Equipotential bonding system

The UFC-V/…-EEx Ultrasonic Flow Converter with cable glands must always be
incorporated into the equipotential bonding system with the UFS 500 F/5STR-EEx 5-beam
Flow Sensor. If the UFC-V/…-EEx is provided with metal rigid conduits, equipotential
bonding is not required but can optional be connected. The bonding conductor must have a
minimum cross-sectional area of 2.5 mm

2

(equivalent to AWG 12) and is provided with a ring

cable lug, which is screwed to the bulge at the base of the flameproof box via a stainless
steel bolt with spring or toothed washer. Make sure that the bolt is tightly fixed.

2.2 Connecting

cables

The following cables are shown in the connection diagram:

Cable A: Signal cable for RS485 communication.
Cable parameters must be in accordance with the regulations in the EN 60079-14 "Electrical
installations in hazardous locations" or an equivalent national standard (e.g. DIN VDE 0165).

Cable B: Power supply cable.
Cable parameters must be in accordance with the regulations in the EN 60079-14 "Electrical
installations in hazardous locations" or an equivalent national standard (e.g. DIN VDE 0165).
Rated voltage:

≥ 500 V

Examples:

H07..-.,

H05..-.


Cable C: Quad coaxial cable.
Type MR04 (supplied by KROHNE).

Technical data:
Test

voltage:

≥ 500 V

Diameter of strand (core and screen):

≥ 0.1 mm

Distributed capacitance (core/screen):

67 pF/m

Distributed inductance (core/screen):

0.4 µH/m


Bonding conductor:
Minimum cross-sectional area:

≥ 2.5 mm

2

(equivalent to AWG 12)