DEVICE DESCRIPTION

2

9

H250 M40

www.krohne.com

01/2012 - 4001485201- MA H250/M40-Ex-II3G-AD R01 en

Using a heat resistant connecting cable

1 Maximum value of the product temperature for the use of a standard connecting cable. For
higher product temperatures a connecting cable with a temperature resistance of 90°C / 194°F is
required.

HT version - high-temperature version with projecting indicator

The permitted ambient temperature range is indicated on the nameplate; depending on the
device version it is T

amb

= -40...+65°C / -40...+149°F or T

amb

= -25...+65°C / -13...+149°F.

The minimum product temperature is -40°C / -40°F.

Temperature table in °C

Heating jacket

T

m

1

None

with

HT-version

T

amb

≤ 40

T

amb

≤ 60

T

amb

≤ 65

DN15, DN25, DN50

DN15, DN25

220

105

75

x

-

175

95

DN80, DN100

DN50, DN80

165

90

75

x

-

155

90

Temperature table in °F

Heating jacket

T

m

1

None

with

HT-version

T

amb

≤ 104

T

amb

≤ 140

T

amb

≤ 149

DN15, DN25, DN50

DN15, DN25

428

221

167

x

-

347

203

DN80, DN100

DN50, DN80

329

194

167

x

-

311

194

AD_H250_M40_II3G_R01_en_4001485201_PRT.book Page 9 Tuesday, January 24, 2012 8:46 AM