Jenway 7305 User Manual
Page 35
8.3.3
Sipper pump
When the sipper is in use the
sipper pump icon is
displayed in the bottom right hand corner of the screen.
The pump direction is displayed by an arrow icon below
the sipper pump icon. The sipper pump can operate
in manual or timed mode, depending on the option
selected in sipper pump settings. To open the sipper
pump settings hold the key below the
sipper pump icon
for 2 seconds.
8.3.3.1 MAnUAL SiPPeR PUMP SettinGS
To operate the sipper pump in manual mode press the key
adjacent to the
manual sipper icon. Select the preferred
pump direction by pressing the key below the
forwards
or
backwards arrow icon. Press the key adjacent to the
tick icon to save and return to the expanded operating
menu.
To perform a measurement place the sipper tubing into
the sample and press the key below the
sipper pump
icon.
Confirmation will be needed to start the sipper pump.
Press the key adjacent to the
tick icon to confirm and
start the sipper pump. Press the key adjacent to the
cross
icon to cancel and return to the expanded operating
menu.
To stop the sipper pump press the key adjacent to the
stop icon. Ensure that the flow through cuvette contains
enough sample before pressing the key below the
measure sample icon.
0.000
100.0
400
09:02
ABS
%T
nm
0.000
100.0
400
09:02
ABS
%T
nm
0.000
100.0
400
09:02
ABS
%T
nm
0.000
100.0
400
09:02
ABS
%T
nm
35