Jenway Genova Nano English User Manual
Page 21

19
9.2
TrOuBLESHOOTING GuIdE
Issue
Solution
Unable to achieve zero absorbance or
Ensure that there is not a sample on the micro volume
100% transmittance when calibrating
accessories read head.
Ensure the instrument lid is closed before and during
the calibration.
Ensure the lamp is working – if the lamp has failed
please request service assistance.
Unable to achieve a reading
Ensure the correct path length is being used.
when measuring a sample
Ensure the sample isn’t too dense so that light is not
transmitted through the sample.
Ensure the lamp is working.