Ht9022, En - 48 – HT instruments HT9022 User Manual
Page 50

HT9022
EN - 48
Range [kW], [kVA]
Resolution [kW], [kVA]
Uncertainty
0.00
÷ 99.99
0.01
±(3.0%rdg+3dgt)
100.0
÷ 999.9
0.1
Uncertainty defined for: sine waveform > 65Hz, Voltage > 10V, Current
≥ 5A,
Pf ≥ 0.5
Active Energy
Range [kWh]
Resolution [kWh]
Uncertainty
0.00
÷ 99.99
0.01
±(2.0%rdg+3dgt)
100.0
÷ 999.9
0.1
Uncertainty defined for: sine wave 10..65Hz, Voltage > 10V, Current
≥ 2A,
Pf ≥ 0.5
Range [kWh]
Resolution [kWh]
Uncertainty
0.00
÷ 99.99
0.01
±(3.0%rdg+3dgt)
100.0
÷ 999.9
0.1
Uncertainty defined for: sine waveform > 65Hz, Voltage > 10V, Current
≥ 5A,
Pf ≥ 0.5
Reactive Power
Range [kVAR]
Resolution [kVAR]
Uncertainty
0.00
÷ 99.99
0.01
±(2.0%rdg+3dgt)
100.0
÷ 999.9
0.1
Uncertainty defined for: sine wave 10..65Hz, Voltage > 10V, Current
≥ 2A,
0.992 ≥ Pf ≥ 0.5
Range [kVAR]
Resolution [kVAR]
Uncertainty
0.00
÷ 99.99
0.01
±(3.0%rdg+3dgt)
100.0
÷ 999.9
0.1
Uncertainty defined for: sine waveform > 65Hz, Voltage > 10V, Current
≥ 5A,
0.992 ≥ Pf ≥ 0.5
Reactive Energy
Range [kVARh]
Resolution [kVARh]
Uncertainty
0.00
÷ 99.99
0.01
±(2.0%rdg+3dgt)
100.0
÷ 999.9
0.1
Uncertainty defined for: sine wave 10..65Hz, Voltage > 10V, Current
≥ 2A,
0.992 ≥ Pf ≥ 0.5
Range [kVARh]
Resolution [kVARh]
Uncertainty
0.00
÷ 99.99
0.01
±(3.0%rdg+3dgt)
100.0
÷ 999.9
0.1
Uncertainty defined for: sine waveform > 65Hz, Voltage > 10V, Current
≥ 5A,
0.992 ≥ Pf ≥ 0.5
Power factor
Range Resolution
Uncertainty
0.20
÷ 1.00
0.01
±3°
Uncertainty defined for: sine waveform 10..65Hz, Voltage > 10V, Current
≥ 2A
Uncertainty defined for: sine waveform > 65Hz, Voltage > 10V, Current
≥ 5A
Voltage and Current harmonics
Harmonic order
Frequency [Hz]
Resolution [V], [A]
Uncertainty
1
ч 25
10
ч 75
0.1
±(5.0%rdg+5dgt)
1
ч 8
75
ч 400
Phase sequence and 1-lead phase coincidence
Range Input
impedance
100
÷ 1000V
1.3M
Ω
Frequency range: 40..70Hz.
(*) Measurement carried out under the following standard conditions: instrument firmly held in the hand, standard shoes, standard floor,
etc.