HT instruments HT9022 User Manual
Page 15

HT9022
EN - 13
5.2.3. Log
By selecting “Log”, the screen to the side will appear.
Press keys F2, F3 (T, S) to change the duration of the
integration period. It will take the following values:
1, 5, 10, 30, 60, 120, 300, 600 or 900 seconds.
Press key F4 (OK) to save the changes made and go back to
the previous screen (see par.5.2).
▼
▲
OK
I n t . P e r i o d :
0 0 1
s
19/01-17:00:00
5.2.4. InRush
By selecting “InRush”, the screen to the side will appear.
Press key F1 (Sel) to move the cursor and keys F2, F3 (T, S)
to change the selected item.
Press key F4 (OK) to save the changes made and go back to
the previous screen (see par.5.2).
• Thresh: threshold value beyond which the inrush current
event is detected and recorded by the instrument. The
current value may be set between 5A and 900A in steps of
1A.
Sel
▼
▲
OK
T h r e s :
0 8 0
W i n d o w : 1 / 1
M o d e : F i x
19/01-17:00:00
• Window: the value of the inrush current measuring window. The following values are
available:
¾
1/1: sampling occurs every half-period;
¾
1/2: sampling occurs one every two half-periods;
¾
1/4: sampling occurs one every four half-periods;
• Mode: the inrush current measuring mode. Following modes are available:
¾
Fix;
¾
Var.
For details about the measurement of Inrush Currents see par. 5.8.2.
5.2.5. Continuity
By selecting “Continuity”, the screen to the side will appear.
Press keys F2, F3 (T, S) to change the setting of the resistance
limit value below which the buzzer will sound. It may be set
between 1
Ω and 150Ω in steps of 1Ω.
Press key F4 (OK) to save the changes made and go back to
the previous screen (see par.5.2).
Sel
▼
▲
OK
R e s L i m :
0 1 0
Ω
19/01-17:00:00