1. SAFETY PRECAUTIONS AND PROCEDURES
This instrument complies with safety IEC/EN61557-1 and IEC/EN61010-1 guidelines
related to electronic measuring instruments.
CAUTION
For your own safety and to avoid damaging the instrument follow the
procedures described in this instruction manual and read carefully all notes
preceded by this symbol
.
When taking measurements:
Avoid doing that in humid or wet places - make sure that humidity is within the limits
indicated in section “environmental conditions”. Avoid doing that in rooms where
explosive gas, combustible gas, steam or excessive dust is present
Keep you insulated from the object under test
Do not touch exposed metal parts such as test lead ends, sockets, fixing objects,
circuits etc
Avoid doing that if you notice anomalous conditions such as breakages, deformations,
fractures, leakages of battery liquid, blind display etc.
Be particularly careful when measuring voltages exceeding 25V in particular places
(building yards, swimming pools, etc.) and 50V in ordinary places to avoid risks of
electrical shocks.
The following symbols are used:
CAUTION - refer to the instruction manual - an improper use may damage the
instrument or its components
DC voltage
AC voltage
Danger high voltage: risk of electric shocks.
Double insulated meter.
1.1. PRELIMINARY
INSTRUCTIONS
This instrument has been designed for use in environments of pollution degree 2
It can be used for tests on electrical installations of overvoltage category III 265V and
550V maximum rated interlinked voltage (and to earth)
You are recommended to respect the usual safety regulations aimed at protecting you
against dangerous currents and protecting the instrument against improper use
Only the original test leads supplied along with the instrument guarantee compliance
with the safety Standards in force. They must be in a good conditions and, if
necessary, replaced with identical ones
Do not test nor connect to any circuit exceeding the specified overload protection
Do not take measurements under environmental conditions exceeding the limits
indicated in this manual
Make sure that batteries are correctly installed
Before connecting the test probes to the installation make sure that the right function is
chosen.