CAUTION

The message on the display means that the instrument is discharging
eventual capacitors. During this phase never disconnect nor touch test leads.

8. At the end of the test the instrument

automatically discharge eventual
capacitors and parasite capacitances
present among the conductors involved
in the measurement

Nominal test voltage
value

9. At the end of the test a screen similar to

this is displayed

Resistance value

10. If the resistance value is higher than the

full scale, a screen similar to this is
displayed.
Note! An insulation value higher than
the full scale is an excellent insulation
value, generally much higher than the
minimum requirements prescribed by
standards

Nominal test voltage
value

11.

+

GO

Press LOCK and GO together to perform the measurement in
continuous mode, the symbol LOCK is displayed. Press GO
another time to stop the continuous mode measurement

12. In lock mode the instrument emits an acoustic signal and the auto power off is

disabled

4.6.1. Measurement time setting mode

Press LOCK more than one
second, a screen similar to this
is displayed

Press the arrow keys to select
the desired time value

Setting values
between 2s an 60s

GO

Press GO to confirm the
selected value

4.6.2. Anomalous

cases

which may occur during M

tests

If, during measurement, the input voltage

present at terminals is higher than 24V,
the instrument suspends the test.
A screen similar to this is displayed
showing the input voltage value.

Input voltage value