beautypg.com

Detcon DM-400IS User Manual

Page 19

background image

RANGE

RANGE

RANGE

RANGE

PPM

0-10ppm

0-25ppm

0-50ppm

0-100ppm

0

40 mV

40 mV

40 mV

40 mV

1

56 mV

46.4 mV

43.2 mV

41.6 mV

5

120 mV

72 mV

56 mV

48 mV

10

200 mV

104 mV

72 mV

56 mV

25

200 mV

120 mV

80 mV

50

200 mV

120 mV

100

200 mV

If calibration is being done with a gas concentration other than 50% of scale, you can determine the proper signal volt-

age by dividing 160 mV by the range of detection and then multiplying that f igure by the concentration of gas. For

example, if you are calibrating a sensor with a range of 0-100ppm with a gas standard that contains 10ppm, you would

divide 160 (mV) by 100 (ppm) and then multiply that f igure by 10 (ppm). Then add the base line of 40 (mV) to arrive

at the proper signal voltage of 56 mV.

3.6.4 Calibration Frequency

In most applications, monthly to quarterly calibration intervals will assure reliable detection. However, industrial envi-

ronments differ. Upon initial installation and commissioning, close frequency tests should be performed, weekly to

monthly. Records should be kept. Less frequent test schedules should be implemented based on analysis of tests prior

to adjustment.

3.7 T

ROUBLE

S

HOOTING

Sensor reads Over-range after Power-up

Probable Cause: Biased sensor requiring additional stabilization time.

1.

Verify if this is a Biased sensor (see section 3.5).

2.

Wait up to 8 hours for unit to come on-scale if using a low range biased sensor.

3.

Verify that there is not large amounts of target gas or interfering gases in background.

Reading Higher than Anticipated

Probable Causes: Target or Interfering gases in background, Incorrect calibration for Zero or Span, Biased sensor still

stabilizing.

1.

Verify no target or interfering gases are present.

2.

Redo Zero and Span calibrations with validated Zero Gas and Span Gas standards.

3.

If recovering after a start-up, give more time to stabilize.

Reading Lower than Anticipated

Probable Causes: Target gas or Interfering gases in background during Zero Calibration, Zero Calibration done before

unit f inished stabilizing, Incorrect Span Calibration.

1.

Redo Zero and Span calibrations with validated Zero Gas and Span Gas standards.

Cannot Zero Calibrate to 40 mV

Probable Causes: Target gas or Interfering gases in background during Zero Calibration, Failed electrochemical sensor.

1.

Verify no target or interfering gases are present.

2.

Redo Zero and Span calibrations with validated Zero Gas and Span Gas standards.

3.

If recovering after a start-up, give more time to stabilize.

Cannot do Span Calibration (not enough adjustment in span pots)

Probable Causes: Failed electrochemical sensor, ice/mud/dust blocking sensor membrane, invalid span calibration gas

do to age and contamination or insuff icient f low rate.

1.

Verify there is no ice/mud/dust blocking sensor membrane.

2.

Redo Span Calibration with validated Span Gas standard (check with Pull Tube).

3.

Replace with new electrochemical sensor.

DM-400IS Toxic Gas Sensors PG.19