Pecifications – Detcon DM-400IS User Manual
Page 13
3.3 S
PECIFICATIONS
Method of Detection
Electrochemical Cell
Electrical Classification
CSA-NRTL (US OSHA) approved* Class 1; Groups B, C, D; Div. 1.
Field Wiring
2 conductor; max 10 ohms single conductor resistance
Input Voltage
11.5-28 VDC
Power Consumption
Normal operation = 4 mA (0.1 watts @ 24VDC); Maximum = 20 mA (0.5 watts @ 24VDC; 0.23 watts @ 11.5VDC)
Output
Linear 4-20 mA DC
Repeatability
± 2% FS
Response
Temperature
Temperature
Humidity
Sensor Cell
Model Number
Gas Name
Time(seconds)
Span Drift
Range °C
Range °F
Range %
Warranty
DM-400IS-C2H3O
Acetyldehyde
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C2H2
Acetylene
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C3H3N
Acrylonitrile
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-NH3 (-20°C)
Ammonia
T90 <60
<1% signal loss/month
-20 to +40
-4 to +104
10 to 95
2 years
DM-401IS-NH3 (-40°C)
Ammonia
T90 <90
<2% signal loss/month
-40 to +40
-40 to +104
5 to 95
2 years
DM-402IS-NH3 (CE)
Ammonia
T90 <90
<2% signal loss/month
-40 to +50
-40 to +122
15 to 90
2 years
DM-400IS-AsH3
Arsine
T90 <60
<5% signal loss/month
-20 to +40
-4 to +104
20 to 95
1 1/2 years
DM-400IS-Br2
Bromine
T90 <60
<2% signal loss/month
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C4H6
Butadiene
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-CS2
Carbon Disulfide
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-CO
Carbon Monoxide
T90
≤
30
<5% signal loss/year
-40 to +50
-40 to +122
15 to 90
3 years
DM-400IS-COS
Carbonyl Sulfide
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-CL2
Chlorine
T90 <60
<2% signal loss/month
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-CLO2 (>10ppm)
Chlorine Dioxide
T90 <60
<2% signal loss/month
-20 to +50
-4 to +122
15 to 90
2 years
DM-401IS-CLO2 (
≤
10ppm)
Chlorine Dioxide
T90 <120
<1% signal loss/month
-20 to +40
-4 to +104
10 to 95
2 years
DM-400IS-B2H6
Diborane
T90 <60
<5% signal loss/month
-20 to +40
-4 to +104
20 to 95
1 1/2 years
DM-400IS-C2H6S
Dimethyl Sulfide
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C3H5OCL
Epichlorohydrin
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C2H5OH
Ethanol
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C2H5SH
Ethyl Mercaptan
T90 <45
<2% signal loss/month
-40 to +50
-40 to +122
15 to 90
2 years
DM-400IS-C2H4
Ethylene
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C2H4O
Ethylene Oxide
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-F2
Fluorine
T90 <80
<5% signal loss/year
-10 to +40
+14 to +104
10 to 95
1 1/2 years
DM-400IS-CH2O
Formaldehyde
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-GeH4
Germane
T90 <60
<1% signal loss/month
-20 to +40
-4 to +104
20 to 95
1 1/2 years
DM-400IS-N2H4
Hydrazine
T90 <120
<5% signal loss/month
-10 to +40
+14 to +104
10 to 95
1 year
DM-400IS-H2 (ppm)
Hydrogen
T90
≤
30
<2% signal loss/month
-20 to +50
-4 to +122
15 to 90
2 years
DM-401IS-H2 (LEL)*
Hydrogen
T90 <60
<2% signal loss/month
-40 to +40
-40 to +104
5 to 95
2 years
DM-400IS-HBr
Hydrogen Bromide
T90 <70
<3% signal loss/month
-20 to +40
-4 to +104
10 to 95
1 1/2 years
DM-400IS-HCL
Hydrogen Chloride
T90 <70
<2% signal loss/month
-20 to +40
-4 to +104
10 to 95
1 1/2 years
DM-400IS-HCN
Hydrogen Cyanide
T90 <40
<5% signal loss/month
-40 to +40
-40 to +104
5 to 95
2 years
DM-400IS-HF
Hydrogen Flouride
T90 <90
<10% signal loss/month
-20 to +35
-4 to +95
10 to 80
1 1/2 years
DM-400IS-H2S
Hydrogen Sulfide
T90
≤
30
<2% signal loss/month
-40 to +50
-40 to +122
15 to 90
2 years
DM-400IS-CH3OH
Methanol
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-CH3SH
Methyl Mercaptan
T90 <45
<2% signal loss/month
-40 to +50
-40 to +122
15 to 90
2 years
DM-400IS-NO
Nitric Oxide
T90
≤
10
<2% signal loss/month
-20 to +50
-4 to +122
15 to 90
3 years
DM-400IS-NO2
Nitrogen Dioxide
T90 <40
<2% signal loss/month
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-O3
Ozone
T90 <120
<1% signal loss/month
-10 to +40
+14 to +104
10 to 95
2 years
DM-400IS-COCL2
Phosgene
T90 <120
<1% signal loss/month
-20 to +40
-4 to +104
10 to 95
1 1/2 years
DM-400IS-PH3
Phosphine
T90 <30
<1% signal loss/month
-20 to +40
-4 to +104
20 to 95
1 1/2 years
DM-400IS-SiH4
Silane
T90 <60
<1% signal loss/month
-20 to +40
-4 to +104
20 to 95
1 1/2 years
DM-400IS-SO2
Sulfur Dioxide
T90
≤
20
<2% signal loss/month
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C4H8S
Tetrahydrothiophene
T90 <30
<2% signal loss/month
-10 to +40
+14 to +104
10 to 95
2 years
DM-400IS-C4H4S
Thiophane
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C6H5CH3
Toluene
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C4H6O2
Vinyl Acetate
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
DM-400IS-C2H3CL
Vinyl Chloride
T90 <140
<5% signal loss/year
-20 to +50
-4 to +122
15 to 90
2 years
* LEL range H2 is not CSA approved.
DM-400IS Toxic Gas Sensors PG.13