beautypg.com

Detcon TP Transmitter Test Fixture User Manual

Page 4

background image

Detcon Model TP Transmitter Test Fixture

Page 4 of 9

3.0 Specifications

Compatible with the Following Models
Model Series TP-424x, TP-524x, TP-550x, TP624x
Input Power
100-240VAC, 50 / 60Hz
Power Consumption
< 17 Watts
Digital Current Meter
Accuracy: ± 0.3% FS
Digital Voltage Meter
Accuracy: ± 0.1% FS
Warranty
Two year
Outputs
Serial RS-485 Modbus™

4.0 Operation

Before using the test fixture, the technician should first be familiar with the operation of
Detcon TP series hydrogen sulfide gas sensors. Review the applicable sensor instruction
manual(s) for detailed operating instructions.

The test fixture is designed to verify performance of the Detcon TP series transmitters
only. The test fixture circuitry mimics the sensor portion of the complete sensor
assembly.

NOTE 1: The test fixture is designed to be used with only one transmitter at a time.
Plugging more than one transmitter in the test fixture at the same time (P1 and P2) will
cause erroneous readings.

NOTE 2: Reference Figure 1 for location of test fixture components. Before starting
verify that SW1 is in the “0” position and that SW2 is in the “NORMAL” position.

4.1 TP-424x Transmitters

Require tools:
Digital Volt-Ohm Meter (DVM)
Jewelers screwdriver

1 - Turn the power switch on. Both digital LED meters should illuminate.

2 - Plug the TP-424x transmitter into P1.