beautypg.com

10 d – Detcon IR-522 User Manual

Page 15

background image

Calibration Fault

If during calibration the sensor circuitry is unable to attain the proper adjustment for zero or span, the sensor will

enter into the calibration fault mode and cause the display to alternate between the sensor’s current status reading and

the calibration fault screen which appears as:

“CAL FAULT.2”.

The following conditions will cause a calibration fault:

1 - Zero calibration cannot converge.

2 - Auto span cannot converge (too noisy or too unstable).

3 - Span gas is not applied before 1 minute elapses.

Fail-Safe/Fault Supervision

Model IR-522 MicroSafe™ sensors are programmed for fail-safe operation. All of the fault conditions listed below

will illuminate the fault LED, cause the mA output to drop to zero (0) mA, and cause the display to read its corre-

sponding fault condition.

Memory Error

If the processor can’t save values to memory, the display will indicate:

“MEMORY ERROR”.

Zero Fault

If the sensor should drift below –10%, the display will indicate:

“ZERO FAULT”.

Lamp Fault

If the lamp signal is lost, the display will indicate:

“LAMP FAULT.2”.

Reference Peak High Fault

If the reference peak signal is too high (>3600), the display will indicate:

“SIGNAL FAULT.31”.

Active Peak High Fault

If the active peak signal is too high (>3600), the display will indicate:

“SIGNAL FAULT.32”.

Reference Peak Low Fault

If the reference peak signal is too low (<500), the display will indicate:

“SIGNAL FAULT.41”.

Active Peak Low Fault

If the active peak signal is too low (<500), the display will indicate:

“SIGNAL FAULT.42”.

Reference Peak to Peak Low Fault

If the reference peak to peak signal is too low (<200), the display will indicate:

“SIGNAL FAULT.51”.

Active Peak to Peak Low Fault

If the active peak to peak signal is too low (<200), the display will indicate:

“SIGNAL FAULT.52”.

3.10 D

ISPLAY

C

ONTRAST

A

DJUST

Model IR-522 MicroSafe™ sensors feature a 16 character backlit liquid crystal display. Like most LCDs, character

contrast can be affected by viewing angle and temperature. Temperature compensation circuitry included in the

MicroSafe™ design will compensate for this characteristic, however temperature extremes may still cause a shift in

the contrast. Display contrast can be adjusted by the user if necessary. However, changing the contrast requires that

the sensor housing be opened, thus declassif ication of the area is required.

To adjust the display contrast, remove the enclosure cover and use a jewelers screwdriver to turn the contrast

adjust screw located beneath the metallic face plate. The adjustment location is marked “CONTRAST”. See f igure 6

for location.

Model IR-522 Combustible Hydrocarbon Sensor PG.15