beautypg.com

Figure #5, Figure #6 – Detcon IR-522 User Manual

Page 12

background image

A magnetic programming tool (see f igure 5) is used to operate the switches. Switch action is def ined as momentary

contact, 3 second hold, and 30 second hold. In momentary contact use, the programming magnet is waved over a

switch location. In 3 second hold, the programming magnet is held in place over a switch location for 3 or more

seconds. In 30 second hold, the programming magnet is held in place over a switch location for 30 or more sec-

onds. Three and thirty second hold is used to enter or exit calibration and program menus while momentary con-

tact is used to make adjustments. The location of “PGM 1” and “PGM 2” are shown in f igure 6.

NOTE: If, after entering the calibration or program menus, there is no interaction with the menu items for more

than 30 seconds, the sensor will return to its normal operating condition.

3.7 C

ALIBRATION

Material Requirements

*

Detcon PN 3270 MicroSafe™ Programming Magnet

*

Detcon PN 6132 Threaded Calibration Adapter

*

Span Gas containing the applicable calibration gas in air. Span gas concentration is recommended at 50% of

range (which is the factory default) at a controlled f low rate of 200 ml/min. Other concentrations can be used

as long as they fall within 10% to 90% of range. See section 3.7.2 for details.

3.7.1 Calibration Procedure - Zero

NOTE: Before performing a zero calibration, be sure there is no background target gas present.

a)

Enter the calibration menu by holding the programming magnet stationary over “PGM 1” (see figure 6) for 3 sec-

onds until the display reads

“1-ZERO 2-SPAN”, then withdraw the magnet. Note that the “CAL” LED is on.

b)

Next, enter the zero menu by holding the magnet stationary over “PGM 1” for 3 seconds until the display

reads:

“ZERO 0%”, then withdraw the magnet. The sensor has now entered the auto zero mode. When it is

complete the display will read

“ZERO COMPLETE” for 5 seconds and then return to the normal operations

menu,

0 % LEL”.

NOTE 1: If the circuitry is unable to adjust the zero to the proper setting the sensor will enter a calibration

fault mode which will cause the display to alternate between the sensor’s current status reading and the calibra-

tion fault screen which appears as:

“CAL FAULT” (see section 3.7.3).

Model IR-522 Combustible Hydrocarbon Sensor PG.12

Magnetic Programming Tool

Figure #5

detcon inc.

Program Switch #2

FLT

CAL

MicroSafe™ LEL Gas Sensor

HOUSTON, TEXAS

PGM

2

PGM

1

MODEL

IR-522

CONTRAST

Fault & Cal LEDs

Program Switch #1

Menu Driven Display

Plug-in Microprocessor Control Circuit

Display Contrast Adjust

Figure #6