Detcon PI-500 User Manual
Page 25
5)
With a f ine tipped tweezers, gcaref illy remove the cell assembly by prying under the cell’s edge where the connector
pins are located.
6)
With a f ine tipped tweezers, grasp the lamp by placing the tips in the housing notch and gently pulling it out. Be
careful not to scratch the lamp or chip the edges.
Cleaning the Lamp
Grab the lamp by the cylindrical glass body and clean the window by rubbiong it against the polishing pad. Use a cir-
cular motion and try to keep the window surface f lat relative to the pad. Five seconds of rubbing should be enough in
most cases. Another indication of cleaning completeness is that about 1/16th of the pads surface is used in the
process.
PI-500 Toxic Gas Sensors PG.25
See also other documents in the category Detcon Sensors:
- TP-700 (48 pages)
- TP-700 SIL 2 Safety Manual (6 pages)
- FP-700 (50 pages)
- FP-700 SIL 2 Safety Manual (6 pages)
- IR-700 (50 pages)
- IR-700 SIL 2 Safety Manual (6 pages)
- DM-700 (58 pages)
- DM-700 SIL 2 Safety Manual (6 pages)
- PI-700 (64 pages)
- PI-700 0-500ppb (59 pages)
- DM-634 (21 pages)
- PI-600 (37 pages)
- DM-600IS (41 pages)
- IR-640 (24 pages)
- IR-622 (24 pages)
- FP-624D-HT (48 pages)
- FP-524HT (22 pages)
- FP-624D (46 pages)
- FP-624C (26 pages)
- TP-624D (44 pages)
- TP-624C (25 pages)
- TP-524D-HRT (52 pages)
- TP-524D (40 pages)
- TP-524C (20 pages)
- FP-524D-HRT (52 pages)
- FP-524D Quick Start Guide (2 pages)
- FP-524D INSTRUCTION MANUAL (43 pages)
- FP-524C (21 pages)
- TP-424C (15 pages)
- DM-200 (50 pages)
- DM-100 (44 pages)
- FP-100 (34 pages)
- SG1 (16 pages)
- DM-534C (22 pages)
- IR-540 (19 pages)
- IR-522 (19 pages)
- DM-434B (11 pages)
- DM-400IS (22 pages)
- FP-424C (15 pages)