Setup menu – DeFelsko PosiTector UTG v.3.0 Quick Guide User Manual
Page 9
Scan
Normally, the PosiTector UTG takes a single spot measurement
at a rate of 6 readings per second while in contact with a surface.
When the probe is lifted, the last reading will remain on the display.
It is sometimes necessary to examine a larger region to locate the
thinnest point. When Scan is selected, the PosiTector UTG will
measure at a rate of 20 readings per second and display min/max
values.
Reset
Reset (soft reset) restores factory settings and returns the
instrument to a known condition. The following occurs:
- All batches, stored measurements, images, and batch names
are erased.
- All calibration adjustments are cleared and returned to the
Gage’s factory calibration settings.
- Menu settings are returned to the following:
Memory = OFF
Statistics Mode = OFF
Hi Lo Alarm = OFF
Scan Mode = OFF
A Scan = OFF
B Scan = OFF
Cal Lock = OFF
Bluetooth = OFF
Perform a more thorough Hard Reset by powering down the
instrument, waiting several seconds, then simultaneously holding
both the center and (+) buttons until the Reset symbol
appears. This returns the instrument to a known, “out-of-the-box”
condition. It performs the same function as a menu Reset with the
addition of:
- Date and Time are not affected by either Reset.
Units = millimeter
Flip Display = Normal
Auto Sync = OFF
White on Black = OFF
Language = English
Battery Type = Alkaline
Backlight = Normal
Bluetooth Streaming = OFF
USB Drive= ON
- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following:
Setup Menu
Setup Menu
NOTE:
Display = None
SE Mode = OFF
8