Scan

Normally, the PosiTector UTG takes a single spot measurement

at a rate of 6 readings per second while in contact with a surface.

When the probe is lifted, the last reading will remain on the display.

It is sometimes necessary to examine a larger region to locate the

thinnest point. When Scan is selected, the PosiTector UTG will

measure at a rate of 20 readings per second and display min/max

values.

Reset

Reset (soft reset) restores factory settings and returns the

instrument to a known condition. The following occurs:
- All batches, stored measurements, images, and batch names

are erased.

- All calibration adjustments are cleared and returned to the

Gage’s factory calibration settings.

- Menu settings are returned to the following:

Memory = OFF

Statistics Mode = OFF

Hi Lo Alarm = OFF

Scan Mode = OFF

A Scan = OFF

B Scan = OFF

Cal Lock = OFF

Bluetooth = OFF

Perform a more thorough Hard Reset by powering down the

instrument, waiting several seconds, then simultaneously holding

both the center and (+) buttons until the Reset symbol

appears. This returns the instrument to a known, “out-of-the-box”

condition. It performs the same function as a menu Reset with the

addition of:

- Date and Time are not affected by either Reset.

Units = millimeter

Flip Display = Normal

Auto Sync = OFF

White on Black = OFF

Language = English

Battery Type = Alkaline

Backlight = Normal

Bluetooth Streaming = OFF

USB Drive= ON

- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following:

Setup Menu

Setup Menu

NOTE:

Display = None

SE Mode = OFF

8