How to Measure

How to Measure

The PosiTector UTGM in multiple-echo mode determines

thickness by measuring the time between at least three consecu-

tive back wall echoes.

steel

Pr

ob

e

1

t

2

t

3

t

1

t

2

t

3

t

Co

at

in

g

Figure 2

Pr

ob

e

In the figure above, multiple-echo mode measures only the time

between echoes. Regardless of whether the steel is coated or not,

all times between echoes are the same. In multiple-echo mode

the Gage determines thickness by measuring t

1

+t

2

+t

3

, dividing it

by six and then multiplying by the velocity of sound for that mate-

rial. The resultant thickness calculation made by the instrument is

therefore an accurate measurement of the steel thickness only,

disregarding the coating thickness.

The velocity of sound is expressed in inches per microsecond or

meters per second. It is different for all materials. For example,

sound travels through steel faster (~0.233 in/µs) than it travels

through plastic (~0.086 in/µs).

1. Remove rubber cap from probe. Couplant (glycol

gel - included) must be applied to the surface to be

tested to eliminate air gaps between the wear face

and the surface. A single drop of couplant is sufficient

when taking a spot measurement.

2. Place the probe flat on the surface. Use moderate pressure to

press against the top of the probe with the thumb or index finger.