Setup menu – DeFelsko PosiTector SPG v.1.1 Quick Guide User Manual
Page 5
5
Selects the type of batteries used in the Gage from a choice of
“Alkaline”, “Lithium” or “NiMH” (Nickel-metal hydride
rechargeable). If NiMH is selected, the Gage will trickle charge the
batteries while connected via USB to a PC or optional AC charger.
The battery state indicator icon is calibrated for the selected
battery type. No damage will occur if the wrong battery type is
selected.
DeFelsko recommends the use of eneloop (NiMH)
rechargeable batteries.
Battery Type
NOTE:
NOTES:
Units = microns
Flip Display = Normal
Auto Sync = OFF
White on Black = OFF
USB Drive= ON
Language = English
Battery Type = Alkaline
Backlight = Normal
Bluetooth Streaming = OFF
- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following:
Perform a more thorough Hard Reset by powering down the
instrument, waiting several seconds, then simultaneously holding
both the center and (+) buttons until the Reset symbol
appears. This returns the instrument to a known, “out-of-the-box”
condition. It performs the same function as a menu Reset with
addition of:
Reset
Reset (soft reset) restores factory settings and returns the
instrument to a known condition. The following occurs:
- All batches, stored measurements, images, and batch names
are erased.
- The zero adjustment is returned to the factory setting. Since
this may not always be a precise zero, the gage should be
checked on the glass plate.
- Menu settings are returned to the following:
Memory = OFF
Statistics Mode = OFF
Hi Lo Alarm = OFF
Bluetooth = OFF
Setup Menu
Setup Menu
Date, Time and WiFi are not affected by either Reset.