background image

4

The factory zero setting may not always be precise,

particularly after a probe tip exchange. For best accuracy, the

gage should be zeroed on the included glass plate.

1.Select from the menu.
2.Press the (+) button to select the number of readings to be

used to obtain an average, typically 3.
3.Repeatedly measure the glass plate. After the last

measurement, the gage will calculate a Zero which represents

the average of all the Zero readings taken.

Glass Plate Zero

Setting the Gage to zero on the glass plate is the preferred zero

adjustment method:

Factory Zero

If a glass plate or suitable smooth, hard surface is not available,

the factory zero setting can be restored:

1. Select from the menu.
2. Press the down button to select “Reset” and press the center

navigation button. xxxx The factory calibration icon iiiii will

appear on the LCD.

Zero

NOTE:

Zero

The PosiTector SPG has only one adjustment point at zero. The

zero point can be set using one of two methods. Both methods are

found under the Zero menu option.

Zero Menu

Zero Menu

Zero

If not, see Returning for Service (pg. 9). Metric example: gage

tolerance is ±5% or ±5 µm whichever is greater. The 76 µm shim

tolerance is ±5 µm. So the average of several measurements

should be between 66 and 86 microns. Inch example: gage

tolerance is ±5% or ±0.2 mils whichever is greater. The 3 mil shim

tolerance is ±0.2 mils. So the average of several measurements

should be between 2.6 and 3.4 mils.