Safety information • 61

English

• Perform function controls regularly as

described in the relevant section in this docu-
ment. Do not operate the device if it is dam-
aged or not working properly.

• Ensure that there is no heat source in the

immediate vicinity. Do not expose to direct
sunlight. The excessive temperature could
damage the electronics.

• Avoid rapid temperature fluctuations. When

the device is transported so that a tempera-
ture difference of more than 20 °C occurs, it
must stay turned off for at least 2 hours before
it can be turned on again. Otherwise, conden-
sation water will form which can damage the
electronics.

• Caustic detergents may damage the surfaces.

Only use a soft cloth dampened with mild
soapsuds to clean the surfaces of the device.

• Only use disinfectants suitable for sensitive

surfaces. Suitable disinfectants are available
from specialist dealers.

Handling measuring

results

WARNING!

Patient hazard

This device is no diagnostic device. It simply
assists the treating physician in establishing a
diagnosis.
– In order to make a precise diagnosis and

initiate therapeutic measures, besides
determination of the weight, further targeted
examinations must be set up by the
physician, and their results must be
considered.

– The responsibility for diagnosis and treatment

lies with the treating physician.