Setting damping – Seca 634 User Manual

Page 22

background image

22

– You can use the arrow keys to adjust

the value in increments according to the

graduations displayed.

– Once the correct value is set, press the

FUNC key again. The pre-Tare function

is now activated.

– Place the chair associated with the

stored tare value on the scale.

– Now place the patient on the scale. The

patient may remain seated.

Read the weighing result off the digital

display.

– To return to normal weighing mode,

press the

FUNC key.

– To switch off the scale, press the Start

key again.

Tip:
Note the weight on your chairs.

Setting damping

Damping is useful for reducing weight determination malfunctions (due to patient move-
ments, for example). The stronger the damping, the more slowly the weight display re-
acts. You can set damping to suit requirements.
– Press the

FUNC key and keep it de-

pressed for 2 seconds.
FIL and the current setting appear in

the display:
0 = slight damping

➞ fast determination of weight

1 = average damping

➞ normal determination of weight

2 = strong damping

➞ slow determination of weight

– Press an arrow key if you want to

change the set value.

– Press the

FUNC key again to store the

displayed value. The scale is then back

in normal weighing mode.

The settings are retained even once the scale is switched off.