Johnson Level & Tool Mfg. 40-6522 User Manual
Page 20
20
©2010 Johnson Level & Tool - Rev. 1
• Once the desired axes has been selected, the UP arrow and
DOWN arrow buttons are used to adjust the position of the
laser plane.
X-Axis Calibration
1. Place the unit into calibration mode as discussed above.
2. Position the unit so that the X-axis is directed to the two targets.
3. Press the button on the remote control to toggle
calibration control to the X-axis.
4. Press the UP arrow and DOWN arrow buttons, respectively to
adjust the laser height to coincide with the zero position of
the target.
5. Press the ENTER button on the remote control to accept the
calibration value, noting that the status indicator goes “OFF”.
6. After calibration is completed, make sure to power off the unit
and then power on again to activate the calibration.
Y-Axis Calibration
1. Place the unit into calibration mode as discussed above.
2. Position the unit so that the Y-axis is directed to the two targets.