Page 4-7 “normal operation conditions, Page 4-8 “emc conformity standards, Page 14-18 “specifications of strain input module – Yokogawa DA100 User Manual
Page 5

IM DA100-01E-88 5/6
for IM DA100-01E 8th Edition
Page 4-5 Changes to the contents of “Information on and Process in Case of Power
Failure”
The DA makes a report immediately after it recovers from the power failure and then stops reporting.
Computing results:
The DA computes data measured up to the point immediately before the power
failure.
Reporting time:
The time when the power failure occurred.
Page 4-7 “Normal Operation Conditions”
Installation category based on IEC61010-1, CSA22.2 No.61010-1
II*
1
Pollution degree based on IEC61010-1, CSA22.2 No.61010-1
2*
2
Warm-up time
At least 30 minutes after power switch-on.
*
1
Describes a number which defines a transient overvoltage condition. It implies the regulation for impulse
withstand voltage. “II” applies to electrical equipment which is supplied from fixed installations like distribution
boards.
*
2
Describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface
resistivity is adhering. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution
occurs.
Page 4-8 “EMC Conformity Standards”
Please refer to these specifications instead of the one printed in the user's manual.
Safety and EMC Standards
CSA
CSA22.2 No.61010-1, installation category II, pollution degree 2
UL
UL61010-1 (CSA NRTL/C)
C-Tick
EN55011 compliance, Class A, Group 1
KC marking
Electromagnetic wave interference prevention standard, electromagnetic wave protection
standard compliance
Page 4-13, 4-17, 4-19, 4-21, 4-22, 4-29 “Specifications of Module”
Installation Category (Overvoltage Category)
CAT II (IEC61010-1, CSA22.2 No.61010-1)
Measurement Category
CAT II (IEC61010-1, CSA22.2 No.61010-1)
Page 14-18 “Specifications of Strain Input Module”
Gauge Method Measurement Range Type Rated Measurement Range
Accuracy
Resolution
Single-gauge
2000µε
-2000 to 2000µε
0.5% of Range
0.1µε
20000µε
-20000 to 20000µε
0.3% of Range
1µε
200000µε
-200000 to 200000µε
0.3% of Range
10µε
Two-gauge
1000µε
-1000 to 1000µε
0.5% of Range
0.1µε
10000µε
-10000 to 10000µε
0.3% of Range
1µε
100000µε
-100000 to 100000µε
0.3% of Range
10µε
Four-gauge
500µε
-500 to 500µε
0.5% of Range
0.1µε
5000µε
-5000 to 5000µε
0.3% of Range
1µε
50000µε
-50000 to 50000µε
0.3% of Range
10µε